Far and away the
most commonly implicated heat sources in OR/surgical fires have been electrocautery
devices (see our December 16, 2014 Patient Safety Tip of the Week “More
on Each Element of the Surgical Fire Triad”). But almost any heat source in
a hyperoxygenated environment can lead to an
OR/surgical fire.
Earlier this year an OR fire related to a cardioversion procedure occurred in Michigan (Counts 2015). Though the hospital apparently did not release details of the incident, news releases (Allen 2015) state that a spark from the cardioverter pad ignited a paper drape and mask covering the patient in an oxygen-rich environment. That resulted in a fire that injured the patient and several staff members.
It’s not the first time that a cardioverter or defibrillator has served as the heat source for an OR fire. In 2012 an OR fire in North Carolina that was fatal to a patient was triggered by a defibrillator (WRAL 2012). And in June 25, 2013 Patient Safety Tip of the Week “Update on Surgical Fires” we discussed a study from a closed claims database (Mehta 2013) in which there was one case where a defibrillator was the heat source.
The 2008 ASA Practice Advisory for the Prevention and Management
of Operating Room Fires (ASA
2008) also mentions defibrillator
paddles or pad as potential heat sources. Other heat sources mentioned
in that advisory include
electrosurgical or electrocautery devices, lasers, heated probes, drills and
burrs, argon beam coagulators, and fiberoptic light
cables.
And in a recent Web
M&M, Mehta and Domino (Mehta
2015) discussed a case where a bag
of bladder irrigating solution hanging over anesthesiologist’s accessory
“slave” monitor dripped onto the power source, producing sparks and flames and
filling the OR with black smoke. While the Mehta case was not an “on-patient”
fire, it illustrates the same OR fire safety principles and the importance of
discussing fire risk during the timeout and doing fire drills that all OR staff
participate in. The staff in that case were very familiar with their roles in
OR fires and responded promptly and appropriately.
AORN (AORN
2015) recently had some useful reminders on prevention of surgical/OR fires
from Mark Bruley of ECRI Institute:
We have long advocated that the surgical fire risk be discussed as part of the pre-op huddle (or pre-op briefing) and, if the case is considered high-risk, respective roles of all OR participants are called out during the surgical timeout. In our January 2011 What's New in the Patient Safety World column “Surgical Fires Not Just in High-Risk Cases” we noted the San Francisco VA checklist “The Surgical Fire Assessment Protocol” (Murphy 2010), This checklist/protocol is actually printed on the reverse side of their larger preoperative checklist. This is really a very good tool! The fire risk is assessed by a simple numerical scale. If the score is 3 (high risk) the rest of the form is filled out, which basically delineates the respective roles of all those participants. That’s a really good way to remind all about their responsibilities if a fire occurred. Bruley, in the AORN article, notes he has seen some surgical fire risk assessments take far too long and recommends teams tailor their fire risk assessment during Time Out to a model created in an algorithm and video from the Anesthesia Patient Safety Foundation (APSF 2010).
We’ve discussed Bruley’s recommendations on oxygen in several of our prior
columns. Perhaps the most important question to ask is “Does the patient need
supplemental oxygen at all?”. But if supplemental
oxygen is required, Bruley notes it is recommended
that it be administered through a secure airway via a tracheal tube or through
a laryngeal mask, except for surgical cases where the patient must be able to
speak during surgery. In such cases, 30% blended oxygen and air should be used
to begin with.
Since the majority
of surgical/OR fires in recent years seem to occur during relatively minor
procedures, many such patients don’t have secure airways in place and then the
need for supplemental oxygen may arise suddenly. The anesthetist then often
simply delivers it via a face mask rather than taking time to use a tracheal
tube or laryngeal mask. Therefore, it remains crucial that in all cases the
surgeon and anesthesiologist communicate clearly, with the surgeon calling out
his/her intention to use the electrocautery device (or other potential heat
source such as a laser) and the anesthesiologist informing the surgeon to wait
until oxygen has been discontinued and allowed to adequately dissipate before
the electrocautery device is used.
In the AORN article Bruley also discusses the importance of educating all team
members about the fire risk and what to do if a fire occurs. While we have
always stressed that each member of the OR team needs to know their specific
role in an OR fire (such as the anesthesiologist turning off the oxygen
supply), Bruley notes that there should not be
specific roles designated for extinguishing a surgical fire because time
is of the essence. He stresses that any member of the surgical team should be
ready to put out a fire.
The recent AHRQ Web
M&M case illustrates the importance of each team member knowing what to do
in the event a fire actually occurs. We can’t overemphasize the importance of training
and drills for OR fires and planning prior to each case. While such fires are
still relatively rare, the consequences can be devastating. The only way to be
prepared for responding to such rare events is to have practiced such responses
frequently.
Our prior columns on surgical fires:
References:
Counts J. U-M cardiovascular center fire started during operation on patient, spread to curtain. Ann Arbor News 2015; July 7, 2015
Allen J. U-M not releasing report on cardiovascular center fire that burned patient. Ann Arbor News 2015; August 7, 2015
http://www.mlive.com/news/ann-arbor/index.ssf/2015/08/u-m_not_releasing_report_of_ca.html
WRAL. Fatal Durham hospital fire may have started during defibrillation. WRAL.com Posted November 6, 2012
http://www.wral.com/explosion-reported-on-sixth-floor-of-durham-regional-hospital/11742138/
Mehta SP, Domino KB. Fire in the Hole!—An OR Fire. AHRQ Web M&M April 2015
https://psnet.ahrq.gov/webmm/case/346/fire-in-the-hole-an-or-fire
ASA (American Society of Anesthesiologists). American Society of Anesthesiologists Task Force on Operating Room Fires. Practice advisory for the prevention and management of operating room fires. Anesthesiology 2008; (108): 786-801
http://anesthesiology.pubs.asahq.org/article.aspx?articleid=1932277
Mehta SP, Bhananker SM, Posner KL, Domino KB. Operating Room Fires: A Closed Claims Analysis. Anesthesiology 2013; 118(5): 1133-1139, May 2013
http://anesthesiology.pubs.asahq.org/article.aspx?articleid=1918584
AORN. 4 actions to improve fire safety. Periop Insider 2015
http://www.informz.net/InformzDataService/OnlineVersion/Pub/bWFpbGluZ0luc3RhbmNlSWQ9MTk3NTM4NA==
Murphy J. A New Effort to Promote Fire Safety in the OR.
Topics In Patient Safety (TIPS) 2010; 10(6): 3
http://www.patientsafety.va.gov/docs/TIPS/TIPS_NovDec10.pdf#page=3
SF VAMC Surgical Fire Risk Assessment Protocol
http://www.patientsafety.va.gov/docs/TIPS/TIPS_NovDec10.pdf#page=3