View as “PDF version”
Patient Safety Tip of the Week
November 19,
2024
Oxygen Disruption
Events
In one of our very first Patient Safety Tips of the Week (March 5, 2007 “Disabled
Alarms”)
we described a “near-miss” in which a patient’s oxygen line had become
disconnected from the wall source of oxygen. No alarm went off, but the patient’s
nurse thought the patient “just didn’t look quite right”. They had been having
trouble with the patient’s pulse oximeter. So, the nurse attached a new pulse
oximeter, and it confirmed low oxygen saturation. Searching for a reason for
the low oxygen saturation, it was discovered that the oxygen line to the
ventilator had become disconnected from the wall source of oxygen. So, the
patient was basically being ventilated with room air. The line was reconnected
to the wall socket and adequate oxygenation rapidly ensured. The patient
suffered no harm due to the incident.
However, one of the first questions raised was “why didn’t an alarm go
off to indicate a low concentration of oxygen coming into the ventilator?” Mechanical
ventilators usually have an oxygen blender alarm. This is an alarm that is
triggered by oxygen concentration falling below the prescribed level. In this particular case, the alarm failed to alert nursing staff
that the oxygen content had dropped.
In
the course of the
investigation, a piece of tape was found covering the oxygen blender alarm. The
alarm, in fact, was functional. It was just not audible because of the tape. We
even considered the worst – had someone intentionally disconnected the oxygen
line and disabled the blender alarm?
One of the nurses had formerly been a respiratory therapist. She noted
that often, during maintenance of the ventilators between use by patients,
staff would test the ventilators with compressed air rather than oxygen (as a
cost-saving measure). They would tape the blender alarm because it would go off
when the compressed air was used instead of oxygen. No other ventilators in use
at this hospital were found to have tape over the blender alarm. However, when
the potential issue was mentioned in an alert to other hospitals in this
multi-hospital system, six other ventilators in actual patient use were found
to have tape over the blender alarm!
The protocol used to assess alarms at set-up time for patient use did not
include a specific assessment of the blender alarm. The lesson learned in this
case is that one should have a checklist-type routine for checking alarms
immediately before a ventilator is used on a patient. And the checklist should
be specific for the ventilator being used. It is not at all uncommon for a
hospital to have multiple different models of ventilators and some may have
alarms that others do not. Therefore, your protocol should include not only
checking those alarms that are on all ventilators but also those that may be
specific to the unit being used. Obviously, those with blender alarms should
have a specific item on the check list to confirm the functionality of that
alarm.
We never did find out how the oxygen line had become
disconnected from the wall outlet (or whether it was never appropriately
connected in the first place.). But, it raises the
question of how often and why oxygen disconnection
events occur. A recent study in the journal Patient Safety (Bocknek
2024) answers some of those questions. Bocknek
et al. analyzed 298 patient safety
event reports in the Pennsylvania Patient Safety Reporting System to understand
the factors contributing to oxygen disruption events.
Almost one quarter of reports involved a patient with high
supplemental oxygen requirements (24.8%). The three care area groups where
oxygen disruption events occurred most frequently were the Emergency Department
(28.9%), Imaging/Diagnostic (17.1%), and Medical/Surgical (12.4%). Over
one-third of all reports occurred on the unit where the patient was admitted (36.6%).
Approximately one quarter of reports occurred upon the patient’s return from
off-unit care (22.1%), and one-fifth of reports occurred upon the patient’s
arrival to off-unit care (17.8%).
They found the most frequent contributing factor to oxygen
disruption was the patient not being transferred to another source of oxygen (45.3%),
followed by tank found empty (35.9%), patient connected to a functioning oxygen
source, no oxygen flowing (8.4%), oxygen delivery device malfunction (7.4%),
and no oxygen available (3.0%).
Roughly 40% of reports involved a hand-off breakdown (41.3%)
and the most frequent breakdowns occurred between a nurse and a patient
transporter (38.2%). Problems maintaining adequate oxygenation are particularly
a problem during transport of patients (within or outside of facilities). One
statistic you’ve often heard us quote is that oxygen therapy has been reported
to be interrupted in as many as 55% of intrahospital transports (PPSA
2005). Failure modes identified in that study included: failure to treat with oxygen when ordered, failure to initiate flow
from the oxygen source, failure to connect the oxygen tubing to the source,
failure to place the oxygen delivery device on the patient, and failure to
anticipate the oxygen demand and provide an adequate supply throughout the
transport. Such interruptions in oxygen delivery were a major reason for
development of “Ticket to Ride” checklists for intrahospital transports. Our
many columns on intrahospital transports and the “Ticket to Ride” concept are
listed below.
Recommendations in the Bocknek
study:
·
Clarify the responsibilities of the entire care
team during hand-offs (e.g., nurses, transporters, procedure technicians, and
therapists)
·
Conduct oxygen rounds in the ED and imaging
·
Evaluate the design of oxygen delivery devices
for usability when procuring new products:
·
A System-Based Solution: Supervisory Control for
Oxygen Tank Monitoring
Interestingly, the Bocknek study found only 7.4% of events classified as “Oxygen delivery device
malfunction”, which described
a break or malfunction involving a device or part of a device that delivers
supplemental oxygen (e.g., wall oxygen connector, oxygen tank, connective
tubing, or ventilator). They don’t have specific recommendations regarding
this category. However, we’d repeat our recommendation about having a check
list of items to verify when setting up a patient on a ventilator or other
oxygen delivery system, including making sure that any monitoring systems (pulse
oximetry, oxygen blender alarms) are functioning properly.
Some of our prior
columns on intrahospital transports and the “Ticket to Ride” concept:
·
April 8, 2008 “Oxygen
as a Medication”
·
November 18, 2008 “Ticket
to Ride: Checklist, Form, or Decision Scorecard?”
·
August 11, 2009
“The
Radiology Suite…Again!”
·
March 13, 2012 “Medical
Emergency Team Calls to Radiology”
·
August
25, 2015 “Checklist for Intrahospital Transport”
·
September
1, 2015 “Smarter Checklists”
·
November
2016 “Oxygen
Tank Monitoring”
·
February
2018 “Oxygen
Cylinders Back in the News”
·
May 22,
2018 “Hazardous
Intrahospital Transport”
·
October
30, 2018 “Interhospital
Transfers”
·
March
31, 2020 “Intrahospital Transport
Issues in Children”
·
June 23,
2020 “Telemetry Incidents”
·
July 14,
2020 “A Thesis on Intrahospital
Transports”
·
August
4, 2020 “Intravenous Issues”
·
February
2021 “Risk from Intrahospital
Transfer: Healthcare-Associated Infection”
·
May 25,
2021 “Yes, Radiologists Have
Handoffs, Too”
·
July
2021 “Unique Way to Rapidly
Identify Oxygen Flow”
·
April
2022 “Safety Issues in
Interhospital Transports”
·
January
31, 2023 “Alert on Oxygen Cylinder Use”
·
June 20, 2023 “What Can Go Wrong on
Intrahospital Transport?”
References:
Bocknek LS, Busog
DNC, Ratwani RM, et al. Unmasking the Contributing
Factors to Oxygen Disruption Events in the Inpatient Environment and Emergency
Department. Patient Safety 2024; 6(1):
Pennsylvania Patient Safety Authority. Patient Safety
Advisory. Continuity of Oxygen Therapy During Intrahospital Transport. PA PSRS
Patient Saf Advis 2005; 2(3): 9-10