The FDA has just released a safety communication regarding surgical fires (FDA 2018). Though it does not contain any new revelations, it provides good recommendations to reduce surgical fires in a concise, practical manner.
It begins with a discussion of the fire triad and
appropriately emphasizes the role of supplemental oxygen. As we have emphasized
in so many of our own columns on surgical fires, it stresses that an open oxygen delivery system (eg.
nasal cannula or mask) presents a greater risk of fire than a closed delivery
system (laryngeal mask or endotracheal tube). It also notes the importance of
draping techniques that avoid accumulation of oxygen in the surgical field.
Weve discussed that surgical fires have in recent years been occurring more often in relatively minor procedures (eg. temporal artery biopsies, plastic procedures or removal of skin lesions on the head/neck). In such cases there may be no need for supplemental oxygen, yet supplemental oxygen is sometimes routinely provided. In others, use of supplemental oxygen is not anticipated but something occurs during the procedure that leads to its use. In both cases, it is critical that there be clear communication and coordination between the anesthesiologist and surgeon regarding cessation of oxygen administration when a heat source is about to be used.
The FDA recommends
that health care professionals and staff who perform surgical procedures be
trained in practices to reduce surgical fires. That training should include
factors that increase the risk of surgical fires, how to manage fires that do
occur, periodic fire drills, how to use carbon dioxide (CO2) fire extinguishers
near or on patients, and evacuation procedures.
The FDA recommends a fire risk
assessment at the beginning of each surgical procedure. We recommend that a fire risk assessment be done both during the
presurgical huddle and as part of the surgical timeout. We continue to
promote use of the SF
VAMC Surgical Fire Risk Assessment Protocol, which can be embedded into
your safe surgery checklist.
The FDA emphasizes
the importance of communication, not only between the anesthesia professional
delivering medical gases and the surgeon controlling the ignition source, but
also amongst the operating room staff applying skin preparation agents and
drapes.
The advisory has
good recommendations regarding surgical suite items that may serve as fuel
sources. It emphasizes the need to allow adequate drying time and prevent
alcohol-based antiseptics from pooling during skin preparation and assess for
pooling or other moisture to ensure dry conditions prior to draping. Were
glad to see they included one of our favorites to avoid in head/neck cases, the 26 ml applicator, in their
recommendations (see our January 10,
2017 Patient Safety Tip of the Week The
26-ml Applicator Strikes Again!). But the FDA also cautions us to be aware of other surgical suite
items that may serve as a fuel source, including products that may trap oxygen,
such as surgical drapes, towels, sponges, and gauze even those which claim to
be "flame-resistant." They also mention patient-related sources such
as hair and gastrointestinal gases. Weve deferred including an article
about a surgical fire related to flatus but, since the FDA mentions it, you can
read it for yourselves (The Asahi Shimbun
2016)!
The section about devices that may serve as an ignition
source is particularly good. First, it advises that alternatives be considered to using an ignition source for surgery of the
head, neck, and upper chest if high concentrations of supplemental oxygen
(greater than 30 percent) are being delivered. As above, if an ignition source
must be used, be aware that it is safer to do so after allowing time for the
oxygen concentration in the room to decrease. It may take several minutes for a
reduction of oxygen concentration in the area even after stopping the gas or
lowering its concentration.
It reminds us to
inspect all instruments for evidence of insulation failure (device, wires, and
connections) prior to use (and do not use if any defects are found). And it
reminds us that, in addition to serving as an ignition source, monopolar energy
use can directly result in unintended patient burns from capacitive coupling
and intra-operative insulation failure. It recommends the following if a
monopolar electrosurgical unit (ESU) is used:
When not in use,
ignition sources, such as ESUs, electrocautery devices, fiber-optic
illumination light sources and lasers should be placed in a designated area
away from the patient (e.g., in a holster or a safety cover) and not directly on the patient or surgical drapes. It also reminds us about
other less common potential ignition sources, such as drills and burrs, argon
beam coagulators, and fiber-optic illuminators.
Lastly, it describes
what to do if a fire occurs:
Surgical fires are devastating and should never occur. Using the precautions noted above and in our multiple columns on surgical fires listed below, you should be able to prevent them. But you must ensure that all staff are educated about surgical fires and do appropriate drills so that everyone knows their role in the unfortunate event that one should occur.
Our prior columns on surgical fires:
References:
FDA (US Food & Drug Administration). Recommendations to Reduce Surgical Fires and Related Patient Injury: FDA Safety Communication. FDA Safety Communication 2018; May 29, 2018
The Asahi Shimbun. Fart blamed for fire during surgery; patient seriously burned. The Asahi Shimbun 2016; October 30, 2016
http://www.asahi.com/ajw/articles/AJ201610300030.html
SF VAMC Surgical Fire Risk Assessment Protocol
https://www.patientsafety.va.gov/docs/TIPS/TIPS_NovDec10.pdf#page=3