In our many columns on patient safety issues in the MRI suite, issues related to implants have often been stressed. Certain implants can become heated during MRI scans, they may become dislodged, or they might malfunction.
Analysis of an FDA database revealed over 600 adverse events related to implantable hearing devices during MRI scanning over a 10-year period (Ward 2020a). Most were cochlear implants, but a few were middle ear implants or bone conduction implants and 2 were brainstem implants. The most common cause of the adverse events was dislocation of the magnet in the implantable device, often causing pain. This often leads to premature cessation of the MRI study. In some cases, manufacturer protocols may allow removal of the magnet prior to the MRI but some events occurred even when following manufacturer guidelines. Headbands or splints are recommended by manufacturers in some cases but, even then, adverse events sometimes occurred. Rarely, there was malfunction of the implanted device. Also, implantable devices may lead to artifacts that make interpretation of certain areas difficult.
Thermal burns are the most common adverse MRI event in an FDA database (Forrest 2020a) but most of these are related to superficial devices like coils and EKG leads. Yet even deep implants that have ferromagnetic properties can heat and cause internal thermal injuries. But reassuring was a recent study showing that clinicians can safely perform clinical 3-tesla MRI scans with metal artifact reduction sequence (MARS) protocols on patients with hip arthroplasty implants without excessive thermal heating of the devices (Forrest 2020b). On the other hand, the thermal effects due to the switching of gradient coil (GC) fields on metallic hip implants has been less well studied (Ward 2020b) For the echo-planar imaging (EPI) sequence, which is needed to perform functional MRI and diffusion-tensor imaging, heating is strongly dependent on the position of the body within the scanner The researchers found almost no heating when the implant is at z ≈ 0 from the scanner center, whereas the worst cases occur with z ≈ 300 mm
All the more important that we always ascertain the presence of any implant in a patient prior to undergoing an MRI scan.
Some of our prior columns on patient safety issues related to MRI:
References:
Ward P. Implantable hearing devices cause MRI safety concerns. AuntMinnieEurope.com 2020; August 10, 2020
https://www.auntminnieeurope.com/index.aspx?sec=sup&sub=mri&pag=dis&ItemID=619122
Forrest W, Thermal burns top FDA list of MRI adverse event reports. AuntMinnie.com 2020; August 13, 2020
https://www.auntminnie.com/index.aspx?sec=rca&sub=ismr_2020&pag=dis&ItemID=129887
Forrest W. ARRS: MRI protocols don't affect hip implant temperature. AuntMinnie.com 2020; May 10, 2019
https://www.auntminnie.com/index.aspx?sec=sup&sub=mri&pag=dis&ItemID=125415
Ward P. Caution needed with MRI of patients with metallic implants. AuntMinnie.com 2020; August 13, 2020
https://www.auntminnie.com/index.aspx?sec=rca&sub=ismr_2020&pag=dis&ItemID=129881
Print PDF version
http://www.patientsafetysolutions.com/
Whats New in
the Patient Safety World Archive