Patient Safety Tip of the Week
Here’s Why
Firefighters Need MRI Training
In many of our columns on MRI safety, we’ve
harped on the need to train your local first responders on responding to
incidents and accidents in MRI centers. A recent article in the firefighting
literature really tells the story why this is so important (DeYear 2025).
Firefighters arrived at the site of a fire
in the early morning hours. There was heavy smoke coming from the one-story structure so they cut and punched through the roof to
ventilate the structure. Officers and firefighters
entered a hallway where smoke was thick and about waist level. They then stepped
into the room they were in front of, searching for the seat of the fire, when
suddenly there was a loud bang. Immediately,
a firefighter yelled “It took it” referring to his Halligan (the axe-like tool
used to breach roofs). The men quickly realized they needed to get out of the
room. The firefighter described the force of the tool being ripped from his
hand “like a scene from Star Wars”. He noted, at one point, he felt as if he
himself was being pulled.

Halligan (left) stuck inside MRI core
\
They
extinguished the fire, but that was not the end of the story. An MRI
maintenance worker and office manager arrived as the companies were mopping up and
the fire investigation was taking place. The maintenance worker informed the
firefighters that he felt confident he could get their tools out of the MRI
machine with the use of a broom handle. Shortly after entering the MRI room,
the maintenance worker let out chilling screams. In his effort to pull one of
the tools loose, the Halligan rotated, and the adz end pinched the maintenance
worker’s fingers and hand to the side of the machine with crushing force. Thinking
quickly, the officers had their crew strip down to their basic clothing before
entering the room. They grabbed an axe handle to help pry
off the tools. The property manager entered the room to help, only to then have
her cell phone sucked into the machine. Firefighters could not move the tool
pinning the maintenance worker’s hand because of the tremendous amount of force
exerted by the MRI machine. The screaming and yelling of the trapped
maintenance worker finally forced them to hit the quench button.
No
one at the scene had ever witnessed what a quench would do. The assumption was
the quench button would discharge helium gas, rendering the magnet to lose its
power. But the discharge of helium occurred with explosive force, being forced
out under tremendous pressure through a pipe to the rooftop. If that pipe had
been clogged, a serious event could have occurred near the machine. Fortunately,
all members were safe and uninjured, and the maintenance worker’s hand was now
free. EMS personnel were able to treat and transport the maintenance worker for
his injury.
The
lessons learned from the incident raised additional concerns and dangers should
be understood by all fire service responders (or any other first responders). DeYear notes that the importance of pre-incident surveys
cannot be overstated. He says “not performing a pre-incident
survey of the MRI facility is a failure on the part of the fire service”. DeYear emphasizes “First-due companies should be familiar with every MRI in their response
area.” MRI units are no longer
found only in hospitals but are also in physicians’ offices, private clinics,
freestanding medical facilities, imaging centers, shopping centers, research
institutes, and even mobile units (several of our prior columns have noted
special difficulties with mobile units).
DeYear goes through
the basics of MRI, the supercooling magnet, types of MRI’s, layout of typical
MRI suites, dangers related to ferromagnetic materials, locations of MRI-compatible
fire extinguishers, signage, location of the quench button, and other key
elements (with good photos and diagrams).
One
important lesson for firefighters is that shutting down power to the building
does not shut down the magnet of the MRI machine. Firefighters often falsely assume
they are safe because the utilities are shut off.
He goes
into detail about what happens and what dangers occur if a quench is initiated,
including the threat of burns, hypothermia, ruptured eardrums, or possible
asphyxiation that can affect anyone in the room as the gas displaces the
oxygen. Visibility can also become a concern. He therefore emphasizes that the
decision to use the quench button should be made with and by staff members of
the facility.
He
then makes the following recommendations for pre-incident surveys and what to
do during an on-site incident:
·
Discover
the boundary of the 5Q line, identifying the invisible magnetic safe zone. 5Q
is the designated “safe area” to operate where the magnetic field of the MRI
machine has dropped below 5 gauss. Operating within
this line poses a danger because of the strength of the machine. This line
varies from machine to machine, as it is determined by the strength of the MRI
magnet. This boundary applies to all directions, including upward and downward.
·
Identify
the location of the emergency shutoff and quench buttons.
·
Generate
a list of names and contact information for those who understand the system and
the dangers.
·
Question
whether the warning signage is appropriate and if it would prove useful in an
IDLH atmosphere where visibility may be nonexistent. Some may even be located
on the floor.
·
Remember
that the magnet is always on, even if the electricity has been shut off. The
magnet will attract and draw anything that is ferrous such as steel, iron,
nickel, and cobalt. The denser the ferromagnetic material is, the higher the
attraction will be.
·
All the
tools we use, along with the personal protective equipment we wear, are a
danger.
·
Flashlights,
portable radios, TICs, and stethoscopes are never allowed within range of the
MRI machine.
·
Personal
items such as cell phones, hair pins, pens, insulin pumps, cochlear implants,
pacemakers, artificial joints, and shrapnel or metal fragments are all dangers
in the MRI room.
·
Confirm
whether there is a nonmagnetic fire extinguisher near the room and where it is
located. It is MRI compatible and often blue or white.
·
If
dangerous or hazardous incidents are identified at the scene of any emergency,
communicate it over the fireground radio immediately.
·
Consider
the high-voltage requirements of the actual machine, especially during fire
control and water usage.
·
Understand
conditions that may require activation of the “emergency power off” or
“emergency shutdown” buttons and where they are located. These buttons may need
to be used for sparks or fire within the computer room and for flooding or
activation of the sprinkler system. Keep in mind that these are different from
the quench button. The MRI machine is still active.
·
Ensure
that you know the circumstances for using and location of the quench button.
·
During
EMS or entrapment situations:
o Follow all the above safety procedures.
o Determine the severity of any injuries.
o Remove the victim before any treatment is
initiated under the guidance of trained MRI personnel.
o Never attempt to resuscitate a victim within
the MRI room.
o During life-or-death situations, the quench
button may be necessary.
o
Always
allow two minutes to pass before entering the area.
That
series of recommendations could be integrated into your training presentations
and manuals for first responders and your own MRI staff. In fact, you could
probably use DeYear’s whole article as the basis for
your training for first responders (with, of course, an on-site tour and
opportunity for two-way discussion).
This
is a really good article that is appropriate for
multiple audiences. Don’t wait for your local fire department (or other first
responders) to reach out to you. Make sure you have contacted them and arranged
for at least an annual inservice on MRI safety (with
on-site visit) and have a mechanism in place for newly added personnel to their
services. The latter is especially important when you deal with volunteer fire
companies.
Some
of our prior columns on patient safety issues related to MRI:
·
February 19, 2008 “MRI Safety”
·
March 17, 2009 “More on MRI Safety”
·
October 2008 “Preventing Infection in MRI”
·
March 2009 “Risk of Burns during MRI
Scans from Transdermal Drug Patches”
·
January
25, 2011 “Procedural
Sedation in Children”
·
February
1, 2011 “MRI
Safety Audit”
·
October
25, 2011 “Renewed
Focus on MRI Safety”
·
August
2012 “Newest
MRI Hazard: Ingested Magnets”
·
October
22, 2013 “How
Safe Is Your Radiology Suite?”
·
October
21, 2014 “The
Fire Department and Your Hospital”
·
August
25, 2015 “Checklist
for Intrahospital Transport”
·
August
2016 “Guideline Update for
Pediatric Sedation”
·
October
2016 “MRI Safety: There’s an App
for That!”
·
January
17, 2017 “Pediatric MRI Safety”
·
August
8, 2017 “Sedation for Pediatric MRI
Rising”
·
March
2018 “MRI Death a Reminder of
Dangers”
·
March
2018 “Cardiac Devices Safe During
MRI But Spinners!?”
·
November
2018 “OMG! Not My iPhone!”
·
April 2,
2019 “Unexpected Events During MRI”
·
September
2019 “New MRI Hazard: Magnetic
Eyelashes”
·
October
15, 2019 “Lots More on MRI Safety”
·
November
5, 2019 “A Near-Fatal MRI Incident”
·
November
2019 “ECRI Institute’s Top 10
Health Technology Hazards for 2020”
·
January
7, 2020 “Even More Concerns About MRI
Safety”
·
March
2020 “Airway Emergencies in the
MRI Suite”
·
October
2020 “New Warnings on Implants and
MRI”
·
January
2021 “New MRI Risk: Face Masks”
·
June 1,
2021 “Stronger Magnets, More MRI
Safety Concerns”
·
November
2021 “Yet Another Risk During MRI”
·
January
2022 “MRI Safety Issues”
·
July 26,
2022 “More Risks in the Radiology
Suite”
·
October 24, 2012 “Serious MRI Accident at
Unregulated MRI Center”
·
November 21, 2023 “Another Terrifying MRI
Accident”
·
January 2024 “Guns and MRI Don’t Mix”
·
February 20, 2024 “What is a “Safety Case”?”
·
June 11, 2024 “Nursing in the MRI Suite”
·
August 13, 2024 “Preventing Thermal Injury
During MRI”
·
October 2024 “Another
MRI Accident Reveals Gap in Oversight
·
October 2024 “MRI
Special Patient Considerations”
·
October 8, 2024 “First Responders and MRI”
·
April 22, 2025 “Updated ACR Manual on MRI
Safety”
·
May 2025 “Add
Another Item to Your MRI Screening Protocols”
References:
DeYear D. May the
Force Be with You: Response to MRI Machines. fireengineering.com
2025; November 1, 2025
https://www.fireengineering.com/firefighting/may-the-force-be-with-you-response-to-mri-machines/
Print “PDF version”

http://www.patientsafetysolutions.com/