Print PDF version
Our November 2025 What's New in the Patient Safety World column MRI Hazards - Big and Small noted a new trend using ferromagnetic nail polish has led to streak artifacts on 3T MRI interfering with scan interpretation (Lou 2025). One example is the "cat eye" manicure -- a pattern achieved by nail technicians swiping a magnet over the magnetic particles in the nail polish. Metallic hair dye may also cause minor distortions on MR images. The article also mentions permanent eyeliner that has ferromagnetic material and many tattoos containing iron oxide as causing not only MRI artifacts but also occasionally causing burns.
MR safety experts Kristin Seitz and Tobias Gilk recently discussed this issue (Seitz 2026). They note there have been lots of warnings on social media about cat-eye nail polish being unsafe to wear during an MRI scan, particularly warnings of the potential for heating and patient burns. That has led to cancellations or delays in patients getting MRI scans in some cases.
They note that metallic particles are not unique to nail polish. Iron oxide is a common ingredient in many cosmetic products -- mascara, bronzer, tinted sunscreen, etc. Yet we dont see the same sorts of warnings about these products. They note that, to date, there have been no FDA MAUDE reported incidences of RF heating of metallic nail polish -- or any other topical cosmetic product. They also note that ASTM/US FDA standards indicate that objects with an electrically conductive length of 2 cm or less are not at risk of clinically significant levels of RF heating in MR imaging at 3.0 T or lower field strengths.
On the other hand, the risks of image distortion or artifacts are real. Yet these artifacts generally only extend only a few inches from the object, so the risk should only really apply to imaging of the hand, wrist or anatomy that is very close to the placement of the hand during imaging.
Seitz and Gilk recommend that, if imaging anatomy other than hands and hands cant be moved >30 cm superior or inferior of image center, such as while imaging the elbow, it is recommended to remove the nail polish. Or, one could move the MRI patients hands toward midline. Also, anatomy further from the walls (i.e., closer to midline of the MR bore) receives significantly less incident RF and will be less likely to heat. Heating risks may also be mitigated through the use of cold compresses.
To avoid artifacts when imaging other body parts (not hands), they recommend simply making sure that the hands with the nail polish arent close to the region of interest.
But, if imaging hands or wrist, it would be advisable to remove the metallic nail polish due to possible artifacts and degradation of image quality.
Seitz also has a video explaining what MRI technologists need to know about imaging and this type of nail polish.
This is very reassuring and hopefully will help avoid delays or cancellations of necessary MRIs in patients having such nail polish.
Some of our prior columns on patient safety issues related to MRI:
·
February 19, 2008 MRI Safety
· March 17, 2009 More on MRI Safety
·
October 2008 Preventing Infection in MRI
· March 2009 Risk of Burns during MRI Scans from Transdermal Drug Patches
·
January
25, 2011 Procedural
Sedation in Children
· February 1, 2011 MRI Safety Audit
· October 25, 2011 Renewed Focus on MRI Safety
· August 2012 Newest MRI Hazard: Ingested Magnets
·
October
22, 2013 How
Safe Is Your Radiology Suite?
·
October
21, 2014 The
Fire Department and Your Hospital
·
August
25, 2015 Checklist
for Intrahospital Transport
·
August
2016 Guideline Update for
Pediatric Sedation
· October 2016 MRI Safety: Theres an App for That!
·
January
17, 2017 Pediatric MRI Safety
·
August
8, 2017 Sedation for Pediatric MRI
Rising
·
March
2018 MRI Death a Reminder of
Dangers
·
March
2018 Cardiac Devices Safe During
MRI But Spinners!?
·
November
2018 OMG! Not My iPhone!
·
April 2,
2019 Unexpected Events During MRI
·
September
2019 New MRI Hazard: Magnetic
Eyelashes
·
October
15, 2019 Lots More on MRI Safety
·
November
5, 2019 A Near-Fatal MRI Incident
·
November
2019 ECRI Institutes Top 10
Health Technology Hazards for 2020
·
January
7, 2020 Even More Concerns About MRI
Safety
·
March
2020 Airway Emergencies in the
MRI Suite
·
October
2020 New Warnings on Implants and
MRI
·
January
2021 New MRI Risk: Face Masks
·
June 1,
2021 Stronger Magnets, More MRI
Safety Concerns
·
November
2021 Yet Another Risk During MRI
·
January
2022 MRI Safety Issues
·
July 26,
2022 More Risks in the Radiology
Suite
· October 24, 2012 Serious MRI Accident at Unregulated MRI Center
· November 21, 2023 Another Terrifying MRI Accident
· January 2024 Guns and MRI Dont Mix
· February 20, 2024 What is a Safety Case?
· June 11, 2024 Nursing in the MRI Suite
· August 13, 2024 Preventing Thermal Injury During MRI
· October 2024 Another MRI Accident Reveals Gap in Oversight
· October 2024 MRI Special Patient Considerations
· October 8, 2024 First Responders and MRI
· April 22, 2025 Updated ACR Manual on MRI Safety
· May 2025 Add Another Item to Your MRI Screening Protocols
· November 2025 MRI Hazards - Big and Small
· January 6, 2026 Heres Why Firefighters Need MRI Training
· March 10, 2026 The Invisible Force Podcast
· May 5, 2026 Sedation in the MRI Suite
· June 2026 Another MRI Risk
References:
Lou N. Magnetic Cosmetics Fly Under the Radar as MRI Risks. MedPage Today 2025; October 16, 2025
https://www.medpagetoday.com/meetingcoverage/asa/117988
Seitz K, Gilk T. Metallic and cat-eye nail polish in MRI. AuntMinnie.com 2026; June 17, 2026
Carey L. MRI safety topic: metallic cat-eye nail polish. AuntMinnie.com 2026; June 17, 2026 (video)
Print PDF version
http://www.patientsafetysolutions.com/
Whats New in
the Patient Safety World Archive