Patient Safety Tip of the Week
Sedation in the
MRI Suite
In last week’s column (April 28, 2026 Patient
Safety Tip of the Week “Patient Safety Events in
Radiology”) we talked about the
many things that can go wrong in the Radiology Suite. Well, the MRI suite is
even more complicated and things can go wrong there, too.
A great example comes from the Finnish Safety
Investigation Authority investigation on a death that occurred during an
MRI procedure in January 2025 (SIAF
2026). A middle-aged
patient was undergoing an MRI for evaluation of shoulder pain at a private
medical facility. An MRI had previously been attempted on the patient without
sedation, but this was unsuccessful. The anesthesiologist suggested performing
an MRI under sedation on the same day because the patient suffered from
claustrophobia. The patient could have waited for an MRI scan performed under
general anesthesia at the University Hospital, for which an orthopedist had
previously provided a referral.
The patient was administered several doses
of anesthetic during the procedure. A total dose was 600 mg of propofol was
given during approximately 20 minutes of anesthesia. The patient also received
2-3 mg of midazolam intravenously at the start and additional later doses of 1
mg. The total dose was 5 mg of midazolam. At the end of the procedure, the
patient received 0.5 mg of flumazenil intravenously to reverse the effect of
the midazolam.
The patient’s condition deteriorated during
the scan and the patient became lifeless. The patient was resuscitated but
later died in a hospital 5 days later.
When the patient was pulled from the MRI
when the scan ended, staff noted the patient’s limbs had become pale and the
patient did not react to movement or speech. The anesthesiologist tried to wake
the patient up and no pulse could be felt. A radiographer went to the recovery
room to get a bag valve mask. The patient was moved onto the floor and pulled into
the control room. A radiographer left to get a defibrillator from the entrance
lobby. Advanced cardiac life support continued with CPR and ventilation of the
patient with a bag valve mask. A radiographer went to the recovery room to get
a resuscitation cart and more supplies. A call button in the imaging room was
used to request additional staff to help with the situation. Several people
arrived on the scene to provide help.
The semi-automatic defibrillator reported
that there was no shockable rhythm. The patient was given two doses of
intravenous resuscitation medication, and a pulse was detected after this. The
patient was connected to a monitor that showed a normal cardiac rhythm. An
attempt was made to insert a breathing tube for the patient using a video
laryngoscope. The resuscitation cart did not contain the correct size of
videolaryngoscope blade, and the first attempt to insert the breathing tube was
unsuccessful. The anesthesiologist then inserted a breathing tube with an
ordinary laryngoscope and continued to assist breathing with the bag valve
mask. The patient’s oxygenation was initially poor, and their blood pressure
was slightly low, but the situation improved to a satisfactory level during
monitoring. The patient began to breathe on their own. The patient was transferred
to a hospital but died 5 days later despite intensive care.
The SIAF investigation found that the amount
of anesthetic medication administered to the patient resulted in a situation
equivalent to general anesthesia rather than sedation. The oxygen flow
administered through an oxygen mask was also too low. An MRI-compatible monitor
required to monitor the patient’s basic vital signs was not available and the
patient’s deteriorating condition was not noticed in time. Apparently, the only
monitoring during the procedure was visually checking for misting of the
patient’s oxygen mask and chest respiration movements (done by the anesthesiologist
in the adjacent room).
In addition, the imaging unit was not
prepared to handle an emergency situation. The resuscitation and patient
transfer equipment had to be retrieved from other spaces in the medical center,
which delayed the start of life support. No regular training for resuscitation
and emergency situations had been organized for imaging unit staff. Oversight
and risk management at the center were also cited as being inadequate, and the
agreement between the center and the company providing anesthesia services was
also general in terms of specifying responsibilities and roles.
SIAF also noted there are no uniform
national guidelines or unambiguous minimum requirements for the implementation
of NORA (non-operating room anesthesia) activities, even though they have
become more common in different environments. This leads to varying practices
and client and patient safety risks.
The SIAF investigation is quite well done.
We encourage you to read the .PDF
of the investigation report. We think there are
valuable lessons learned here and that all out-of-hospital MRI facilities should
be asking whether similar incidents could happen at their facilities.
Lastly, there does appear to be at least in
Finland some regulatory oversight of MRI units. You’ll recall from the recent Invisible Force Podcast
segments and several of our prior columns on MRI safety that such oversight is
glaringly missing in the US.
Some
of our prior columns on patient safety issues in the radiology suite:
·
October 16, 2007 “Radiology
as a Site at High-Risk for Medication Errors”
·
February 19, 2008 “MRI
Safety”
·
September 16, 2008 “More
on Radiology as a High Risk Area”
·
October 7, 2008 “Lessons
from Falls....from Rehab Medicine”
·
October 2008 “Preventing
Infection in MRI”
·
March 17, 2009 “More
on MRI Safety”
·
March 2009 “Risk
of Burns during MRI Scans from Transdermal Drug Patches”
·
August 11, 2009
“The
Radiology Suite…Again!”
·
January
2010 “Falls in the Radiology Suite”
·
August 2010 “Sedation
Costs for Pediatric MRI”
·
January
25, 2011 “Procedural Sedation in Children”
·
February
1, 2011 “MRI Safety Audit”
·
October
25, 2011 “Renewed Focus on MRI Safety”
·
March 13, 2012 “Medical
Emergency Team Calls to Radiology”
·
August
2012 “Newest MRI Hazard: Ingested Magnets”
·
October
22, 2013 “How Safe Is Your Radiology Suite?”
·
February
25, 2014 “Joint Commission Revised Diagnostic Imaging
Requirements”
·
July
2014 “New MRI Risks: for Staff!”
·
July 1,
2014 “Interruptions and Radiologists”
·
November
2014 “More Radiologist Interruptions”
·
October
21, 2014 “The Fire Department and Your Hospital”
·
June 23,
2015 “Again! Mistaking Antiseptic Solution for
Radiographic Contrast”
·
August
25, 2015 “Checklist for Intrahospital Transport”
·
March
22, 2016 “Radiology
Communication Errors May Surprise You”
·
August
2016 “Guideline
Update for Pediatric Sedation”
·
October
2016 “MRI
Safety: There’s an App for That!”
·
January
17, ,2017 “Pediatric
MRI Safety”
·
August
8, 2017 “Sedation
for Pediatric MRI Rising”
·
November
14, 2017 “Tracking
C. diff to a CT Scanner”
·
March
2018 “MRI
Death a Reminder of Dangers”
·
March
2018 “Cardiac
Devices Safe During MRI But Spinners!?”
·
April
2018 “Radiologists
Get Fatigued, Too”
·
May 2018
“Cost
of Interrupting a Radiologist”
·
November
2018 “OMG!
Not My iPhone!”
·
December
11, 2018 “Another NMBA Accident”
·
April 2,
2019 “Unexpected Events During MRI”
·
September
2019 “New MRI Hazard: Magnetic
Eyelashes”
·
October
15, 2019 “Lots More on MRI Safety”
·
November
5, 2019 “A Near-Fatal MRI Incident”
·
November
12, 2019 “Patient Photographs Again
Help Radiologists”
·
November
26, 2019 “Pennsylvania Law on
Notifying Patients of Test Results”
·
January
7, 2020 “Even More Concerns About MRI
Safety”
·
July 14,
2020 “A Thesis on Intrahospital
Transports”
·
August
25, 2020 “The Off-Hours Effect in
Radiology”
·
October 2020 “New
Warnings on Implants and MRI”
·
January 2021 “New
MRI Risk: Face Masks”
·
May 25, 2021 “Yes, Radiologists Have
Handoffs, Too”
·
June 1,
2021 “Stronger Magnets, More MRI
Safety Concerns”
·
November
2021 “Yet Another Risk During MRI”
·
January
2022 “MRI
Safety Issues”
·
February
2022 “Managing Incidental Findings”
·
June
2022 “Where Are You Barcoding?”
·
July 12,
2022 “Radiologists Racked by
Interruptions”
·
July 26,
2022 “More Risks in the Radiology
Suite”
·
May 2023
“Factors
Associated with Radiology Diagnostic Errors”
·
September 12, 2023 “Radiology and
Hospital-Acquired Infections”
·
October 24, 2012 “Serious MRI Accident at
Unregulated MRI Center”
·
November 21, 2023 “Another Terrifying MRI
Accident”
·
January 2024 “Guns
and MRI Don’t Mix”
·
June 11, 2024 “Nursing in the MRI Suite”
·
August 13, 2024 “Preventing Thermal Injury
During MRI”
·
October 2024 “Another
MRI Accident Reveals Gap in Oversight
·
October 2024 “MRI
Special Patient Considerations”
·
October 8, 2024 “First Responders and MRI”
·
April 22, 2025 “Updated ACR Manual on MRI
Safety”
·
July 8, 2025 “Applying Human Factors
Concepts to Radiology”
·
July 22, 2025 “What the Heck are LLM’s?”
·
November 2025 “MRI
Hazards - Big and Small”
·
January 6, 2026 “Here’s Why Firefighters Need
MRI Training”
·
March 10, 2026 “The Invisible Force Podcast”
·
April 28, 2026 “Patient Safety Events in
Radiology”
Some
of our prior columns on patient safety issues related to MRI:
·
February 19, 2008 “MRI Safety”
·
March 17, 2009 “More on MRI Safety”
·
October 2008 “Preventing Infection in MRI”
·
March 2009 “Risk of Burns during MRI
Scans from Transdermal Drug Patches”
·
January
25, 2011 “Procedural
Sedation in Children”
·
February
1, 2011 “MRI
Safety Audit”
·
October
25, 2011 “Renewed
Focus on MRI Safety”
·
August
2012 “Newest
MRI Hazard: Ingested Magnets”
·
October
22, 2013 “How
Safe Is Your Radiology Suite?”
·
October
21, 2014 “The
Fire Department and Your Hospital”
·
August
25, 2015 “Checklist
for Intrahospital Transport”
·
August
2016 “Guideline Update for
Pediatric Sedation”
·
October
2016 “MRI Safety: There’s an App
for That!”
·
January
17, 2017 “Pediatric MRI Safety”
·
August
8, 2017 “Sedation for Pediatric MRI
Rising”
·
March
2018 “MRI Death a Reminder of
Dangers”
·
March
2018 “Cardiac Devices Safe During
MRI But Spinners!?”
·
November
2018 “OMG! Not My iPhone!”
·
April 2,
2019 “Unexpected Events During MRI”
·
September
2019 “New MRI Hazard: Magnetic
Eyelashes”
·
October
15, 2019 “Lots More on MRI Safety”
·
November
5, 2019 “A Near-Fatal MRI Incident”
·
November
2019 “ECRI Institute’s Top 10
Health Technology Hazards for 2020”
·
January
7, 2020 “Even More Concerns About MRI
Safety”
·
March
2020 “Airway Emergencies in the
MRI Suite”
·
October
2020 “New Warnings on Implants and
MRI”
·
January
2021 “New MRI Risk: Face Masks”
·
June 1,
2021 “Stronger Magnets, More MRI
Safety Concerns”
·
November
2021 “Yet Another Risk During MRI”
·
January
2022 “MRI Safety Issues”
·
July 26,
2022 “More Risks in the Radiology
Suite”
·
October 24, 2012 “Serious MRI Accident at
Unregulated MRI Center”
·
November 21, 2023 “Another Terrifying MRI
Accident”
·
January 2024 “Guns and MRI Don’t Mix”
·
February 20, 2024 “What is a “Safety Case”?”
·
June 11, 2024 “Nursing in the MRI Suite”
·
August 13, 2024 “Preventing Thermal Injury
During MRI”
·
October 2024 “Another
MRI Accident Reveals Gap in Oversight
·
October 2024 “MRI
Special Patient Considerations”
·
October 8, 2024 “First Responders and MRI”
·
April 22, 2025 “Updated ACR Manual on MRI
Safety”
·
May 2025 “Add
Another Item to Your MRI Screening Protocols”
·
November 2025 “MRI
Hazards - Big and Small”
·
January 6, 2026 “Here’s Why Firefighters Need
MRI Training”
·
March 10, 2026 “The Invisible Force Podcast”
References:
SIAF (Safety Investigation Authority,
Finland). A patient’s death as a result of anaesthesia administered during an
MRI procedure highlights the difficulty of monitoring, self-supervision and
implementing non-operating room anaesthesia (NORA). Safety Investigation Authority,
Finland 2026;
SIAF (Safety Investigation Authority,
Finland). Patient’s death as a result of anaesthesia administered during
magnetic resonance imaging in January 2025. Safety Investigation Authority,
Finland 2026;